8508A
Users Manual
resistance from the instrument's INPUT Hi and Lo terminals and the
resulting potential difference is sensed by the
SENSE Hi and Lo terminals
 To cancel autoranging and enter a manually selected range, press any range
selection or the Auto softkey. The Multimeter enters the selected range or
reverts to the auto-selected range.
TRU
Configuration (Resolution, Fast, and Low Current)
Press the
Config key to see the TRU CONFIG menu.
adj082f.eps
TRU CONFIG Menu
 Provides selection of the resolution for subsequent readings.
 Select a higher read rate. This will result in an increase in uncertainty due to
noise.
 Controls the measurement currents. The current in use for the selected range is
indicated in the left-hand display together with the maximum voltage that will
appear across the measured resistance at full scale.
TRU
Resolution
Press the Resl key to see the TRU RESL menu:
adj083f.eps
TRU RESL Menu
5
6
7
8
5-1/2-digits resolution
6-1/2-digits resolution
7-1/2-digits resolution
8-1/2-digits resolution
Refer to Table 3-2 for the ADC integration times applicable to the various resolution and
Fast On/Off settings.
To revert to the
TRU CONFIG menu, press the Config key.
3-16