6
Adjustment and Verification
Performance Verification
Table 6-10. Resistance Verification Points and Sequence
8508A Range Resistance Tolerance
1
Normal and Tru
2
Normal LoI and Tru LoI
2
2 0 Perform input zero operation to calibrator zero output
1 ± 0.0000240 ± 0.0000240
20 0
Perform input zero operation to calibrator zero output
10 ± 0.000133 ± 0.000133
200 0 Perform input zero operation to calibrator zero output
100 ± 0.00101 ± 0.00113
2 k 0 k
Perform input zero operation to calibrator zero output
1 k ± 0.0000101 k ± 0.0000113 k
20 k 0 k
Perform input zero operation to calibrator zero output
10 k ± 0.000101 k ± 0.000113 k
200 k 0 k Perform input zero operation to calibrator zero output
100 k ± 0.00101 k ± 0.00107 k
2 M 0 M
Perform input zero operation to calibrator zero output
1 M ± 0.0000117 M ± 0.0000142 M
20 M 0 M
Perform input zero operation to calibrator zero output
10 M ± 0.000320 M ± 0.000570 M
200 M 0 M Perform input zero operation to calibrator zero output
100 M ± 0.01950 M ± 0.18700 M
2 G 0 G
Perform input zero operation to calibrator zero output
1 G ± 0.0018750 G ± 0.0018750 G
6-45