EasyManua.ls Logo

Fluke 8558A - AC Voltage Calibration Data (cont.); AC Voltage kHz and MHz Ranges

Fluke 8558A
46 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Reference Multimeter / 8½ Digit Multimeter
Calibration Verification
17
Range Frequency
RMS
Filter
Reference
Value
Permissible Error (±)
Fluke
Calibration
Uncertainty (±)
8588A 8558A
100 mV 2 kHz 10 Hz 100.000 0 mV 0.009 5 mV 0.018 1 mV 0.002 4 mV
100 mV 10 kHz 10 Hz 100.000 0 mV 0.014 8 mV 0.018 1 mV 0.002 4 mV
100 mV 30 kHz 10 Hz 100.000 0 mV 0.014 9 mV 0.032 3 mV 0.002 4 mV
100 mV 60 kHz 10 Hz 100.000 0 mV 0.072 3 mV 0.098 0 mV 0.002 4 mV
100 mV 100 kHz 10 Hz 100.000 0 mV 0.072 3 mV 0.098 0 mV 0.002 5 mV
100 mV 300 kHz 10 Hz 100.000 0 mV 0.300 0 mV 0.405 0 mV 0.005 8 mV
100 mV 1 MHz 10 Hz 100.000 0 mV 1.420 0 mV 2.070 0 mV 0.040 0 mV
100 mV 2 MHz 10 Hz 100.000 0 mV 2.590 0 mV 2.970 0 mV 0.140 0 mV
100 mV 4 MHz 10 Hz 100.000 0 mV 6.450 0 mV 7.490 0mV 0.140 0 mV
100 mV 6 MHz 10 Hz 100.000 0 mV 11.590 0 mV 13.150 0mV 0.150 0 mV
100 mV 8 MHz 10 Hz 100.000 0 mV 11.590 1 mV 13.150 0mV 0.150 0 mV
100 mV 10 MHz 10 Hz 100.000 0 mV 20.690 0 mV 24.150 0mV 0.150 0 mV
1 V DC 10 Hz 0.000 000 V 0.000 080 V 0.000 083 V 0.000 013 0 V
1 V DC 10 Hz 1.000 000 V 0.000 210 V 0.000 249 V 0.000 017 0 V
1 V DC 10 Hz -1.000 000 V 0.000 210 V 0.000 249 V 0.000 017 0 V
1 V 10 Hz 10 Hz 1.000 000 V 0.000 090 V 0.000 129 V 0.000 028 0 V
1 V 40 Hz 10 Hz 1.000 000 V 0.000 090 V 0.000 129 V 0.000 019 0 V
1 V 105 Hz 10 Hz 1.000 000 V 0.000 090 V 0.000 129 V 0.000 019 0 V
1 V 1 kHz 10 Hz 0.010 000 V 0.000 010 V 0.000 014 V 0.000 001 2 V
1 V 1 kHz 10 Hz 0.100 000 V 0.000 010 V 0.000 025 V 0.000 002 1 V
1 V 1 kHz 10 Hz 0.200 000 V 0.000 020 V 0.000 036 V 0.000 004 0 V
1 V 1 kHz 10 Hz 0.400 000 V 0.000 040 V 0.000 059 V 0.000 008 0 V
1 V 1 kHz 10 Hz 0.600 000 V 0.000 060 V 0.000 083 V 0.000 012 0 V
1 V 1 kHz 10 Hz 0.800 000 V 0.000 070 V 0.000 106 V 0.000 016 0 V
1 V 1 kHz 10 Hz 1.000 000 V 0.000 090 V 0.000 129 V 0.000 018 0 V
1 V (AC) 1 kHz 10 Hz 1.000 000 V 0.000 090 V 0.000 129 V 0.000 018 0 V
1 V 1 kHz 10 Hz 1.200 000 V 0.000 110 V 0.000 152 V 0.000 023 0 V
1 V 2 kHz 10 Hz 1.000 000 V 0.000 090 V 0.000 129 V 0.000 018 0 V
1 V 10 kHz 10 Hz 1.000 000 V 0.000 150 V 0.000 180 V 0.000 018 0 V
1 V 30 kHz 10 Hz 1.000 000 V 0.000 280 V 0.000 320 V 0.000 018 0 V