8588A/8558A
Service Manual
18
Range Frequency
RMS 
Filter
Reference 
Value
Permissible Error (±)
Fluke 
Calibration
Uncertainty (±)
8588A 8558A
1 V 60 kHz 10 Hz 1.000 000 V 0.000 720 V 0.000 980 V 0.000 018 0 V
1 V 100 kHz 10 Hz 1.000 000 V 0.000 720 V 0.000 980 V 0.000 020 0 V
1 V 300 kHz 10 Hz 1.000 000 V 0.003 100 V 0.004 050 V 0.000 051 0 V
1 V 1 MHz 10 Hz 1.000 000 V 0.014 300 V 0.020 700 V 0.000 300 0 V
1 V 2 MHz 10 Hz 1.000 000 V 0.025 900 V 0.029 700 V 0.001 400 0 V
1 V 4 MHz 10 Hz 1.000 000 V 0.064 500 V 0.074 900 V 0.001 400 0 V
1 V 6 MHz 10 Hz 1.000 000 V 0.115 900 V 0.131 500 V 0.001 400 0 V
1 V 8 MHz 10 Hz 1.000 000 V 0.115 900 V 0.131 500 V 0.001 500 0 V
1 V 10 MHz 10 Hz 1.000 000 V 0.206 900 V 0.241 500 V 0.001 500 0 V
10 V DC 10 Hz 0.000 00 V 0.000 59 V 0.000 65 V 0.000 150 V
10 V DC 10 Hz 10.000 00 V 0.001 91 V 0.002 31 V 0.000 190 V
10 V DC 10 Hz -10.000 00 V 0.001 91 V 0.002 31 V 0.000 190 V
10 V 10 Hz 10 Hz 10.000 00 V 0.000 90 V 0.001 29 V 0.000 300 V
10 V 40 Hz 10 Hz 10.000 00 V 0.000 89 V 0.001 29 V 0.000 170 V
10 V 105 Hz 10 Hz 10.000 00 V 0.000 89 V 0.001 29 V 0.000 170 V
10 V 1 kHz 10 Hz 1.000 00 V 0.000 15 V 0.002 46 V 0.000 018 V
10 V 1 kHz 10 Hz 10.000 00 V 0.000 89 V 0.001 29 V 0.000 170 V
10 V (AC) 1 kHz 10 Hz 10.000 00 V 0.000 89 V 0.001 29 V 0.000 170 V
10 V 1 kHz 10 Hz 12.000 00 V 0.001 06 V 0.001 52 V 0.000 220 V
10 V 2 kHz 10 Hz 10.000 00 V 0.000 89 V 0.001 29 V 0.000 170 V
10 V 10 kHz 10 Hz 10.000 00 V 0.001 49 V 0.001 81 V 0.000 170 V
10 V 30 kHz 10 Hz 10.000 00 V 0.002 84 V 0.003 23 V 0.000 170 V
10 V 60 kHz 10 Hz 10.000 00 V 0.007 23 V 0.009 80 V 0.000 170 V
10 V 100 kHz 10 Hz 10.000 00 V 0.007 23 V 0.009 80 V 0.000 170 V
10 V 300 kHz 10 Hz 10.000 00 V 0.031 00 V 0.040 50 V 0.000 290 V
10 V 1 MHz 10 Hz 10.000 00 V 0.143 00 V 0.207 00 V 0.001 500 V
10 V 2 MHz 10 Hz 3.000 00 V 0.123 20 V 0.152 10 V 0.004 500 V
10 V 4 MHz 10 Hz 3.000 00 V 0.645 00 V 0.749 00 V 0.004 500 V
10 V 6 MHz 10 Hz 3.000 00 V 1.159 00 V 1.315 00 V 0.004 400 V