Reference Multimeter / 8½ Digit Multimeter
Calibration Verification
19
Range Frequency
RMS
Filter
Reference
Value
Permissible Error (±)
Fluke
Calibration
Uncertainty (±)
8588A 8558A
10 V 8 MHz 10 Hz 3.000 00 V 1.159 00 V 1.315 00 V 0.004 400 V
10 V 10 MHz 10 Hz 3.000 00 V 2.069 00 V 2.415 00 V 0.004 600 V
100 V DC 10 Hz 0.000 0 V 0.005 7 V 0.006 3 V 0.001 50 V
100 V DC 10 Hz 100.000 0 V 0.019 7 V 0.022 9 V 0.001 90 V
100 V DC 10 Hz -100.000 0 V 0.019 7 V 0.022 9 V 0.001 90 V
100 V 10 Hz 10 Hz 100.000 0 V 0.009 6 V 0.012 9 V 0.002 60 V
100 V 40 Hz 10 Hz 100.000 0 V 0.009 6 V 0.012 9 V 0.002 10 V
100 V 105 Hz 10 Hz 100.000 0 V 0.009 6 V 0.012 9 V 0.002 10 V
100 V 1 kHz 10 Hz 10.000 0 V 0.001 5 V 0.002 5 V 0.000 22 V
100 V 1 kHz 10 Hz 100.000 0 V 0.009 6 V 0.012 9 V 0.002 10 V
100 V (AC) 1 kHz 10 Hz 100.000 0 V 0.009 6 V 0.012 9 V 0.002 10 V
100 V 1 kHz 10 Hz 120.000 0 V 0.011 4 V 0.015 2 V 0.002 50 V
100 V 2 kHz 10 Hz 100.000 0 V 0.009 6 V 0.012 9 V 0.002 10 V
100 V 10 kHz 10 Hz 100.000 0 V 0.012 2 V 0.018 1 V 0.002 10 V
100 V 30 kHz 10 Hz 100.000 0 V 0.028 4 V 0.032 3 V 0.002 10 V
100 V 60 kHz 10 Hz 100.000 0 V 0.072 3 V 0.098 0 V 0.002 40 V
100 V 100 kHz 10 Hz 100.000 0 V 0.072 3 V 0.098 0 V 0.003 30 V
100 V 200 kHz 10 Hz 100.000 0 V 0.510 0 V 0.670 00 V 0.014 00 V
100 V 300 kHz 10 Hz 22.000 0 V 0.159 0 V 0.248 80 V 0.003 30 V
100 V 1 MHz 10 Hz 22.000 0 V 0.933 8 V 1.326 80 V 0.005 90 V
1 kV DC 10 Hz 0.000 V 0.082 02 V 0.089 02 V 0.015 0 V
1 kV DC 10 Hz 1000.000 V 0.248 02 V 0.320 02 V 0.019 0 V
1 kV DC 10 Hz -1000.000 V 0.248 02 V 0.320 02 V 0.019 0 V
1 kV 40 Hz 10 Hz 1000.000 V 0.148 00 V 0.220 00 V 0.021 0 V
1 kV 105 Hz 10 Hz 1000.000 V 0.148 00 V 0.220 00 V 0.021 0 V
1 kV 1 kHz 10 Hz 100.000 V 0.043 60 V 0.057 10 V 0.002 1 V
1 kV 1 kHz 10 Hz 500.000 V 0.090 00 V 0.129 50 V 0.010 0 V
1 kV 1 kHz 10 Hz 1000.000 V 0.148 00 V 0.220 00 V 0.021 0 V
1 kV (AC) 1 kHz 10 Hz 1000.000 V 0.148 00 V 0.220 00 V 0.021 0 V
1 kV 2 kHz 10 Hz 1000.000 V 0.148 V 0.220 V 0.021 0 V