EasyManua.ls Logo

Fluke SimpliFiber Pro - Evaluating Reference Value Goodness; Testing Cords in Both Directions

Fluke SimpliFiber Pro
76 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SimpliFiber Pro Meter and Test Kits
Users Manual
48
2 Set the Reference
2-1
Set the reference as described on page 26.
2-2
Make sure that the reference value is good:
For 50/125 µm fiber, the reference value must be
better than -24.50 dBm (for example, -23.50 dBm is a
better value)
For 62.5/125 µm fiber, the reference value must be
better than -20.00 dBm (for example, -19.50 dBm is a
better value)
For 9/125 µm fiber, the reference value must be
better than -8.00 dBm (for example, -7.50 dBm is a
better value)
If the value is not good, do the reference procedure
again with a different test reference cord.
3 Test the Cord in Both Directions
3-1
Disconnect the test reference cord from the meter
(Figure A-3,
A
).
3-2
Connect the cord you want test (Figure A-3,
B
).
3-3
On the meter, press M until LOSS appears.
3-4
If the source’s AUTO LED is not on, press A.
3-5
If you want the source to automatically switch
wavelengths, press
Z until the wavelength LEDs
blink alternately. Or you may press Z to switch
wavelengths as needed.
Note
If you are not using a SimpliFiber Pro source, set
the source to the desired wavelength and to
continuous wave output. On the meter, press S V
to select the wavelength generated by the source.
3-6
Compare the loss measurement to these limits:
Multimode test reference cord: 0.10 dB or less
Singlemode test reference cord: 0.20 dB or less
If the connectors show no contamination or damage,
but the loss is higher than these limits, the cord does
not comply with ISO/IEC 14763-3. Do not use the cord
as a test reference cord.
3-7
Connect the cord in the opposite direction (Figure A-3,
C
), then repeat step 3-6.

Other manuals for Fluke SimpliFiber Pro

Related product manuals