Fig. 96 Define export folder
The last step is to select a path for the exported files.
Click on the question mark to get help.
Fig. 97
This refers to the empty check sample cell which is filled with a local sample and
used together with prediction models suitable for that sample type. For information
about the artificial FOSS Check Sample, see section 4.5.2 FOSS Check Sample.
A check sample (or control sample) is measured on a regular basis to validate the
stability of the analyser and everything around it that affects its performance
(environment etc.). The test frequency is determined by your local quality routines,
but a daily check sample test is the normal recommendation.
The local check sample is a real sample and as such it will not last forever. It will
dry out or in other ways change its properties over time. Therefore, the check
sample needs to be replaced on a regular basis and this frequency can either be
determined in advance (based on experience with your samples) or monitored in
ISIscan Nova by monitoring how the check sample results change.
Replacing the check sample with a new sample will most likely change the
measured result(s) for it; therefore the targets for the check sample test must be
changed accordingly. This can be done manually in FossManager or via a new
check sample definition in ISIscan Nova.
Before running the check sample you must create a check sample Product in
FossManager.
1. Import the prediction model(s) you intend to use for your check sample and
link a Parameter profile. (Create a new Parameter profile if needed.)