116 FONIX 7000 Hearing Aid Analyzer
OSPL90 Sweep Meas Delay—The delay between frequencies in the
OSPL90 sweep.
Res Crv Sweep Start Delay—The delay after the first tone is presented in
the response curve (at 50 or 60 dB SPL) sweep and before the first mea
-
surement is taken.
Res Crv Sweep Meas Delay –The delay between frequencies in the
response curve (at 50 or 60 dB SPL) sweep.
Misc Start Delay—The delay after the first tone is presented and before
the first measurement is taken in the HFA/SPA, reference test gain, equiv
-
alent input noise, and harmonic distortion measurements.
Misc Meas Delay—The delay between frequencies in the measurements
mentioned above.
Note: ANSI 87 combines the “OSPL90” and “Res Crv” delays into the “Sweep
Start” delay settings.
AGC Settings
The purpose of the following settings is identical for each I/O frequency.
AGC frequency—Status of the I/O test at the selected frequency. Turn the
test ON or OFF.
I/O Start—The delay after the first tone is presented in the I/O sweep and
before the first measurement is taken.
I/O Meas—The delay between frequencies in the I/O sweep.
Attack Window—The amount of time the analyzer measures the attack
time of the hearing aid. Select a time that is at least twice as long as the
attack time specified by the manufacturer.
Release Window—The amount of time the analyzer measures the release
time of the hearing aid. Select a time that is at least twice as long as the
release time specified by the manufacturer.
Note: ANSI 87 only measured I/O at 2000 Hz, and the Attack and Release win
-
dow settings are not available.
Misc Settings
Print Label—Status of the printing label. Choose ON or OFF.
Printer—Printer used for printouts. Choose INTERNAL or EXTERNAL.
See Section 2.4.