EasyManua.ls Logo

Frye FONIX 7000 - Running a Single Frequency Measurement; Running a Three-Frequency Average

Frye FONIX 7000
216 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Basic Sound Chamber Tests 73
3.4.5 Running a Single Frequency Measurement
At times, you may want to find out what the frequency response of the hearing
aid is to a single pure-tone signal. See Figure 3.4.5.
1. Press [MENU] from the Coupler screen.
2. Use [∨, ∧] to select Static Tone under Source Settings.
3. Use [<, >] to choose SINGLE.
4. Press [EXIT] to return to the Coupler screen. The single tone will automati
-
cally be running.
5. Adjust the frequency using the [<, >] buttons. Adjust the amplitude using
the [
∧,∨] buttons.
Figure 3.4.5—Measuring a single frequency
3.4.6 Running a Three-Frequency Average
The ANSI standard includes some three frequency average tests. You can run
a three frequency average test (apart from the test sequence) in the Coupler
screen. See Figure 3.4.6. To do this:
1. Press [MENU] from the Coupler screen.
2. Use [∨, ∧] to select Static Tone under Source Settings.

Table of Contents