Automated Test Sequences 123
Misc. Start Delay—The delay after the first tone is presented and before
the first measurement is taken in the HFA/SPA, reference test gain, equiv
-
alent input noise, and harmonic distortion measurements.
Misc. Meas. Delay—The delay between frequencies in the measurements
mentioned above.
I/O Start Delay—The delay after the first tone is presented in the I/O
sweep and before the first measurement is taken.
I/O Meas. Delay—The delay between frequencies in the I/O sweep.
Misc. Settings
Print Label—Status of the printing label. Choose ON or OFF.
Printer—Printer used for printouts. Choose INTERNAL or EXTERNAL.
See Section 2.4.
5.4 JIS
The JIS 2000 automated test sequence allows you to testing hearing aids accord-
ing to the Japanese Instrument Standard. Press [F7] from the Opening screen to
enter the JIS test screen.
5.4.1 Running JIS
Follow these instructions to run a JIS automated test sequence.
1. Press [F7] from the Opening screen to enter the JIS test screen.
2. Use [F1] to select the ear.
3. Use [F2] to select the aid type: Press [F2], use [∨, ] to make your selection,
and [>] to complete the selection and close the pop-up menu.
4. Use [F3] to select the amplitude of the full-on gain measurement.
5. Use [F4] to select the test frequency that you want to use.
6. Use [F5] to select the type of coupler you are using for the test. The 2-cc
selection is used for the HA-1 and HA-2 coupler. The MZ selection is used
for the MZ couplers.
7. Press [MENU] and use the arrow keys to make any desired changes to the
menu items. See Section 5.4.4 for more details. Press [EXIT] to close the
local menu.
8. Press [START] to start the test sequence. The instrument will run a measure
-
ment and then pause.
9. Turn the gain of the aid down to the reference test position. Match the target
value within 1 dB.
10. Press [START] to complete the test sequence.
11. Press [PRINT] to print test results.