EasyManua.ls Logo

Frye FONIX 7000 - Directional Hearing Aids; Using the Insertion Gain Technique

Frye FONIX 7000
216 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
160 FONIX 7000 Hearing Aid Analyzer
5. Press [START] to measure the frequency response of the hearing aid. Press
[STOP] when the measurement has stabilized.
6. Use [F2] to select the REAR 3.
7. Use [F5] to select Composite.
8. Use [∨, ∧] to set the source amplitude to 65 dB SPL.
9. Press [START] to start the measurement. After the measurement has sta
-
bilized, press [STOP] to stop the measurement. Figure 6.5.4 illustrates an
example of a digital hearing aid with noise suppresion. Curve 2 shows the
response of the aid to the Digital Speech curve, and Curve 3 shows the
response of the aid to the Composite signal. .
Figure 6.5.4—Testing noise suppression of a digital hearing aid. REAR 2 was tested using
Digital Speech. REAR 3 was tested using the Composite Signal.
6.6 Directional Hearing Aids
The directional advantage provided by a hearing aid can be easily measured
with real-ear measurements. This section describes a method of measuring
directionality using the Real-ear Insertion Gain screen.
6.6.1 Using the insertion gain technique
The Real-ear Insertion Gain screen is ideal for performing real-ear measure-
ments of hearing aid directionality.
To test directionality:
1. Set up the patient for real-ear measurements by inserting the probe micro-
phone and hearing aid.
2. Place the sound field speaker behind the patient at 180° or 135° azimuth.
Point the reference microphone of the probe microphone set towards the
sound field speaker.
3. Enter the Real-ear Insertion Gain screen.

Table of Contents