EasyManua.ls Logo

Frye FONIX 7000 - Page 169

Frye FONIX 7000
216 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Real-Ear Measurements 161
4. Press [LEVEL] to level the sound field speaker.
5. Use [F2] to select REUG 1. (Press [F2] to open the pop-up selection menu.
Use [∨, ∧] to select REUG 1. Press [>] to complete the selection and close
the menu.)
6. Use [F5] to set the source type to DIG SPEECH.
7. Use [∨, ∧] to set the source amplitude to 50 dB SPL.
8. Press [START] to start the frequency response measurement. When the mea
-
surement has stabilized, press [STOP]. This is the “reverse” measurement
curve.
9 Use [F2] to select REAG 2.
10. Use [F5] to set the source type to DIG SPEECH.
11. Use [∨, ] to set the source amplitude to 50 dB SPL.
Figure 6.6.1—Testing directionality. REUR 1 was tested with the source at 135° azimuth. REAR
2 was tested with the source at 45° azimuth. Curve 6 illustrates the directional advantage of the
hearing aid.
12. Reposition the patient so that the sound field speaker is placed at 0° or 45°
azimuth. The reference microphone should be repositioned so that it is fac
-
ing towards the sound field speaker.
13 Press [LEVEL] to re-level the sound field speaker. This will temporarily clear
(but not erase) the reverse measurement.

Table of Contents