EasyManua.ls Logo

Frye FONIX 7000 - Testing with Bias Signals

Frye FONIX 7000
216 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Basic Sound Chamber Tests 85
5. Use [F5] to choose the DIG SPEECH signal source. It will now have the
ICRA speech weighting. See Figure 3.5.2 for a comparison of an aid tested
with the ANSI and ICRA speech weightings.
5. Use [F5] to choose the DIG SPEECH signal source. It will now have the
ICRA speech weighting. See Figure 3.5.2 for a comparison of an aid tested
with the ANSI and ICRA speech weightings.
Figure 3.5.2—Comparison of a digital aid tested with Digital Speech ICRA (CRV1), Digital
Speech ANSI (CRV2), and the Composite signal (CRV3).
3.5.3 Testing with Bias Signals
You can add a pure-tone bias signal to the Digital Speech signal, creating a con-
tinuous signal at the frequency of your choice. This will allow you to see how
the aid reacts to the presence of noise at different frequencies. To add a bias sig
-
nal:
1. Select the DIG SPEECH signal with [F5] in the Coupler screen.
2. Press [MENU] to enter the local menu.
3. Use [∨, ∧] to select Bias Tone under Source Settings. Use the [START] key to
skip ahead sections, if necessary.
4. Use [<, >] to select the bias amplitude.
5. Press [EXIT] to return to the Coupler screen.
6. Press [START] to start the frequency response measurement. The Digital
Speech signal will include a bias signal at the amplitude you specified in
Step 4.

Table of Contents