EasyManua.ls Logo

Frye FONIX 7000 - Page 144

Frye FONIX 7000
216 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
136 FONIX 7000 Hearing Aid Analyzer
4. Make sure HTL is selected above F2. You should see a cursor in the HTL
column for the selected ear in the middle of the display. If the cursor is in a
different column, press [F2] to select HTL.
5. Use the arrow keys to input the client’s HTL (hearing threshold level) val
-
ues. Use [<, >] to input the amplitude. Use [∨, ∧] to select the frequency.
6. Use [F5] to select the client’s age. (Press [F5]. Use [∨, ] to make the selec
-
tion. Press [>] to complete the selection and close the pop-up window.)
7. Press [F2] to select UCL if you would like to input your client’s measured
uncomfortable levels. Press [F3] if you would like to predict the UCL values.
(Pascoe [1988] is used for calculating UCL values.)
8. Press [F4] and select COPY EAR with [∨, ] if you would like to copy the
data from the selected ear into the other ear. Press [>] to make the selection.
This is useful if your client has a similar loss in both ears.
Figure 6.3.1—Audiogram Entry screen
To specify the transducer used to take the audiometric measurements,
press [MENU] and use the arrow keys to specify the Aud. Transducer under
Targets.

Table of Contents