EasyManua.ls Logo

Fuji Electric ZRE-2 - Page 21

Fuji Electric ZRE-2
85 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
14
INZ-TN2ZRE-E

)
Input signal:
External zirconia O
2
analyzer : Zirconia O
2
sensor signal (Fuji ZFK7 output)
External O
2
analyzer : 0 to 1 V DC (DC input resistor of 1MΩ or more)
It is used when the external zirconia O
2
analyzer or the external O
2
analyzer is specied as
order.
Connect the dedicated connector (accessory) to the output of the external Zirconia analyzer or
the external O
2
analyzer prepared separately.
In case of an external O
2
analyzer, input a signal of 0 to 1 V DC with respect to O
2
full scale
of the analyzer. The O
2
concentration display, output, and O
2
correction can be performed.
Do not connect when the built-in O
2
analyzer is installed.
O
2
sensor input is not isolated. It is recommended to isolate when an external O
2
analyzer is
installed apart from this analyzer. Zirconia O
2
sensor (Fuji ZFK7) should be installed at a loca-
tion that is as close to this instrument as possible.
* How to connect the O
2
signal to the dedicated connector (accessory).
AI
-
AI+
< External input > A/I connector (O2 sensor input)
To O2 sensor
Turn
Solder the signal cable to the terminal.
+

Table of Contents

Related product manuals