SS1807-N002 Checkup, Replacement and Adjustment - 27/137
1.1.9 Air Guide Pipe
■ Removal Procedures
(1) Remove light source unit following the procedures (1) to (6) of “1.1.8 Light Source Unit”.
(2) Remove [1] screws x 6, then remove [2] LG cover assy.
Key Name ID Q’ty Note
1 Screw 308S0426* 6 120Ncm
2 LG Cover Assy Not Supply Parts 1
(3) Remove [3] screws x 2, then remove [4] air tube assy.
Key Name ID Q’ty Note
3 Screw 308S0422* 2 90Ncm
4 Air Tube Assy Not Supply Parts 1