SS1807-N002 Checkup, Replacement and Adjustment - 88/137
1.1.29 Harness ELC-OP
■ Removal Procedures
(1) Remove the front cover assy following the procedures (1) to (3) of “1.1.3 Front Cover”.
(2) Remove the connector of [1] pump assy from CN11 of ELC PCB.
(3) Remove [2] special screw x 4, then [1] pump assy shift forward.
Key Name ID Q’ty Note
1 Pump Assy 133Y200006* 1
2 Special Screw 308N120037* 4 90Ncm
(2) Remove [4] screw x 2 and [5] clamp x 2.
Key Name ID Q’ty Note
4 Screw 308S0406* 2 120Ncm
5 Clamp 316S1189* 2
[1]
CN11
[2]
[5]
[5]
[4]
[4]