Chapter 2 Installation Notes on Installation Site
Chapter 2 Installation
This section describes ECG installation, cable connection, power ON/OFF, and paper replacement procedures.
Notes on Installation Site
Take the following precautions when selecting the installation location.
Radio interference occurs if there are strong radiated radio waves nearby (leakage currents,
static induction and electromagnetic induction enter by the path shown in the figure).
Select a location where no high-voltage power cables or power lines with a high load will
pass by the device or the patient's bed.
Equipment such as X-ray devices, ultrasound devices, radio devices, stands and fluorescent
lighting also causes interference.
Select a location where the room temperature will remain at 20°C to 25°C.
Select a location with low humidity.
Install the device where there is enough space between the wall to connect/disconnect the
power supply cable.
If any abnormality is found on the device, immediately turn OFF the power, and disconnect
the power supply cable from the outlet.
Do not install the device in a location where it is difficult to disconnect the power supply
cable.
Install this device in a place where power supply cable can be easily disconnected.