BEFORE TESTING
Calibration
You must calibrate the meter every time you begin to use a new vial
of ß-Ketone test strips by setting the meter with the correct code. Test
results may be inaccurate if the code number displayed on the
monitor does not match the number printed on the strip vial.
How to Code Your Meter (for ß-Ketone)(b)
1. Insert the code strip when the monitor is off. Wait until the
code number appears on the display.
NOTICE:
Make sure the code number on display, code strip, and test strip vial
are the same. The code strip should be within the expiry date;
otherwise, an error message may appear.
2. Remove the code strip, the display will show “OK”. This tells you
that the meter has finished coding and is ready for ß-ketone testing.
Checking the Code Number
You need to make sure that the code number displayed on the meter
matches the number on your test strip vial before you proceed. If it
matches, you can proceed with your test. If the codes do not match,
please stop testing and repeat the calibration procedure. If the
problem persists, contact Customer Service for help.
NOTICE:
The codes used in this manual are examples only; your meter
may display a different code.
7