SPECIFICATIONS
Model No: Care Touch
Dimension: 96.5 (L) x 60.5 (W) x 22 (H) mm
Weight: 67g
Power Source: Two 1.5 V AAA alkaline batteries
Display: LCD
Memory: 1000 measurement results with respective date and time
External output: Strip port cable communication
Auto electrode insertion detection
Auto reaction time count-down
Auto switch-off after 3 minutes without action
Temperature Warning
Operating Condition: 50°F ~ 104°F (10°C ~ 40°C), 10% to 85%
R.H. (non-condensing)
Meter Storage/Transportation Conditions: -4°F ~ 140°F (-20°C ~
60°C), below 95% R.H.
Glucose Test Strip Storage / Transportation Conditions: 35.6°F to
86.0°F (2°C to 30°C), 10% to 85% R.H. (non-condensing), up to
21 months for un-opened vial
ß-Ketone Test Strip Storage / Transportation Conditions: 35.6°F to
86.0°F (2°C to 30°C), 10 to 85% R.H. (non-condensing), up to
18 months for un-opened vial
Glucose Measurement Units: Either mg/dL or mmol/L
Glucose Measurement Range: 20 to 600 mg/dL (1.1 ~ 33.3 mmol/L)
ß-ketone Measurement Units: mmol/L
ß-ketone Measurement Range: 0.1 ~ 8.0 mmol/L
This device has been tested to meet the electrical and safety
requirements of: IEC/EN 61010-1, IEC/EN 61010-2-101,
IEC/EN 61326-1, IEC/EN 61326-2-6.