Error Messages
MESSAGE
WHAT IT MEANS
ACTION
Appears when the
Replace the battery
E-b
battery cannot provide
immediately.
enough power for a test.
E-2
Expired code strip.
Repeat the test with a new lot
of test strip.
E-U
Appears when a used
Repeat the test with a new test
test strip is inserted.
strip.
E-0
Repeat the test with a new test
strip. If the meter still does not
E-A
Problem in operation.
work, please contact the
customer service for
E-E
assistance.
Appears when test strip
E-F
is removed while
Repeat the test with a new test
counting down, or other
strip.
problems in operation.
Appears when the wrong
Make sure to insert the right
code strip for the meter.
E-C
code strip is inserted or
Check if the code numbers on
other coding errors.
the code strip and the strip vial
label are the same.
Appears when ambient
System operation range is
46.4°F ~ 113°F (8°C ~ 45°C).
E-t
temperature is below or
Repeat the test after the meter
above system operation
range.
and test strip are in the above
temperature range.