To obtain disinfecting wipes and other information, please contact
Future Diagnostics LLC at 1.866.890.8500. You can also purchase it
at Amazon (www.amazon.com) or Walmart (www.walmart.com).
Disinfecting Procedures
1. Wipe all meter’s exterior surface display and buttons by three
passes vertically and three passes horizontally with a folded
disinfecting towelette. Hold the meter with the test strip slot pointing
down and wipe the area around the test slot but be careful not to
allow excess liquid to get inside. Keep meter wet with disinfection
solution contained in the wipe for a minimum of 2 minutes for Micro-
Kill+™ wipes. (o)
2. Remove the wipe. Allow the meter surface to dry completely.
3. Discard the used wipes and never reuse them. Users should
wash hands thoroughly with soap and water after handling the
meter, lancing device, or test strips.
Improper system cleaning and disinfection may result in meter
malfunction. If you have a question, please contact local
customer service at 1.866.890.8500 for assistance.
This device has been validated to withstand up to 10950 cleaning and
disinfection cycles using the recommended disinfecting wipe/towelette.
The tested number of cycles is estimated by 10 cleaning and
disinfection cycles per day over 3 years. The meter should be
replaced after the validated number of cleaning and disinfection cycles
or the warranty period, whichever comes first.
Stop using the meter if you see any signs of deterioration. For example:
Meter cannot be turned on
LCD display cracks or becomes cloudy
Buttons no longer function
Meter outer casing cracks
Data cannot be transmitted to pc
18