WARNING:
It is important to make sure that the LCD displayed code is the
same as the code on your test strip vial before testing. Failure to do
so will get inaccurate results.
The code number for ß-Ketone strip is three-digits; please
ensure you are using the correct strips for the test.
If the LCD displayed code is not the same as the code on your
test strip vial and the code number cannot be updated, please
contact Customer Service for assistance.
Control Solution Testing
Our Control Solution contains a known amount of glucose that
reacts with test strips and is used to ensure your meter and test
strips are working together correctly.
Do a control solution test when:
You first receive the meter
At least once a week to routinely check the meter and test strips
You begin using a new vial of test strips
You suspect the meter or test strips are not working properly
Your blood glucose or ketone test results are not consistent
with how you feel, or if you think the results are not accurate
Practicing the testing process, or
You have dropped or think you may have damaged the meter
Performing a Control Solution Test
To perform a control solution test, you will need: (c), (d) and (e).
1. Insert the test strip into the meter. There will be a strip light
indicating the test strip is inserted into the test slot. Wait for the
meter to display “ ” and “ ”.