2. Press M to mark this test as a control solution test
With “QC” displayed, the meter will store your test result in
memory under “QC”. If you press M again, the “QC” will disappear
and this test is no longer a control solution test.
3. Apply Control Solution (h)
Shake the control solution vial thoroughly before use. Squeeze out
a drop and wipe it off, then squeeze out another drop and place it
on the tip of the vial cap. Hold the meter to move the absorbent
hole of the test strip to touch the drop. Once the confirmation
window fills completely, the meter will begin counting down. To
avoid contaminating the control solution, do not directly apply
control solution onto a strip.
4. Read and Compare the Result
After counting down to 0, the control solution test result will appear
on the display. Compare this result with the range printed on the
test strip vial and it should fall within this range. If not, please read
the instructions again and repeat the control solution test.
With “QC” displayed, the meter will store your test result in
memory under “QC”.
Out-of-range results
If you continue to have test results fall outside the range printed on the test
strip vial, the meter and strips may not be working properly. Do NOT test
your blood. Call customer service at 1.866.890.8500 for help.
NOTE:
The control solution range printed on the test strip vial is for
control solution use only. It is not a recommended range for your
blood glucose or ketone level.
See MAINTENANCE section for important information about
your control solutions.
9