Hex
command
Description Data eld Response Data/Response description
Sets the calibration
factor that modies
pressure.
33 SET AUTO-RESTART
Sets supply to
automatically
restart on power
up.
"YES" or
"NO"
34 GET AUTO RESTART
Reads auto restart
status of supplies.
"YES" or
"NO"
37 START PUMP
Enables high
voltage.
None or
"1"
38 STOP PUMP
Disables high
voltage.
None or
"1"
3C GET SETPOINT
Reads conguration
of the set point.
None or
"1"
N, E, X.XE-XX,
Y.YE-YY, O
N = Setpoint Number
E = Enabled (1 = yes, 0 = no)
X.XE-XX = On Pressure
Y.YE-YY = O Pressure
O = Setpoint On (1 = yes, 0 =
no)
3D SET SETPOINT
Congures a
specied set point.
N, S,
X.XE-
XX,
Y.YE-
YY
N = Setpoint Number
S = Supply (0 = Inactive, 1 =
Supply 1) X.XE-XX = On
Pressure
Y.YE-YY = O Pressure
44 LOCK KEYPAD
Established Remote
Mode and locks the
front panel except
HV o and power
keys.
45 UNLOCK KEYPAD
Unlocks all front
panel keys.
04/2021 - ©Gamma
Page 45900026_F
900026_F - Serial (J3) and ethernet (J2) operation