Hex
command
Description Data eld Response Data/Response description
50 GET ANALOG MODE
Reads the current/
pressure analog
output mode.
0-10
(7 is MPC
only)
0 = Logarithmic pressure 1 =
Logarithmic current
2 = Volts per 1.0uA 3 = Volts
per 10.0uA
4 = Volts per 100.0uA 5 =
Volts per 1.0mA
6 = Volts per 10.0mA 8 =
Volts per 1.0 nA
9 = Volts per 10.0 nA
10 = Volts per 100.0 nA
51 SET ANALOG MODE
Sets the current/
pressure analog
output mode.
n n = 0-10, 7 is MPC only 0 =
Logarithmic pressure 1 =
Logarithmic current 2 = Volts
per 1.0uA
3 = Volts per 10.0uA 4 =
Volts per 100.0uA 5 = Volts
per 1.0mA
6 = Volts per 10.0mA 8 =
Volts per 1.0 nA
9 = Volts per 10.0 nA
10 = Volts per 100.0 nA
61 IS HIGH VOLTAGE
ON
Indicates if the high
voltage enabled
None or
"1"
"YES" or
"NO"
62 SET SERIAL
ADDRESS
Sets the controllers
serial address.
nnn nnn = new serial address
(000-255)
68 SET HV
AUTORECOVERY
Sets the power auto
recovery mode. For
Auto HV restart, if
HV is on and power
interrupted, unit
will power up and
start HV. For auto
power, if HV was on
and power
interrupted, unit
will power up only
(does not start HV).
n n = 0-2
0 = disabled
1 = enable auto HV restart
2 = enable auto power
recovery
04/2021 - ©Gamma
Page 46900026_F