EasyManua.ls Logo

Gamma Vacuum SPCe - Page 47

Gamma Vacuum SPCe
60 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Hex
command
Description Data eld Response Data/Response description
69 GET HV
AUTORECOVERY
Displays the mode
of Auto Recovery.
n n = 0-2
0 = disabled
1 = enable auto HV restart
2 = enable auto power
recovery
8F SET_FIRMWARE_UPDATE
Sets the SPCe to
ash load mode for
rmware updates.
D3 SET COMM MODE
Sets the
communication
mode to local, full,
or remote.
N N is the mode 0 = Local
1 = Remote
2 = Full
D4 GET COMM MODE
Returns the current
communication
mode.
N N is the mode 0 = Local
1 = Remote
2 = Full
46 GET/SET SERIAL
COMM
Gets/Sets baud rate,
parity, data bits,
and stop bits. If no
parameters are
specied, the
current values are
returned.
None or
"B, P, D, S"
(set mode)
None or "B,
P, D, S"
(get mode)
B = baud rate
P = parity ("n", "e", "o") D =
data bits ("7", "8")
S = stop bits ("1", "2")
47 GET/SET ETHERNET
IP
Gets/Sets Ethernet
IP address. If no
parameters are
specied, the
current values are
returned.
None or
"X.X.X.X"
(set mode)
None or
"X.X.X.X" (get
mode)
48 GET/SET ETHERNET
MASK
Gets/Sets Ethernet
mask. If no
parameters are
specied, the
current values are
returned.
None or
"X.X.X.X"
(set mode)
None or
"X.X.X.X" (get
mode)
04/2021 - ©Gamma
Page 47900026_F
900026_F - Serial (J3) and ethernet (J2) operation

Table of Contents

Other manuals for Gamma Vacuum SPCe

Related product manuals