RC1xx LED Lighting Controllers - User Manual
32
Where m = ags as follows:
bit 1 = 0
1
E ag set (error detection enabled)
E ag cleared (error detection disabled)
bit 2 = 0
1
P ag set (positive triggers)
P ag cleared (negative triggers)
bit 3 = 0
1
Autosense set (controller waits for light to be
connected
No Autosense (controller assumes light is
connected)
bit 6 = 1 Enable current adjust
Disable current adjust
Note: Bit 0 is the least signicant bit.
Set the internal trigger
These commands control the internal trigger. When set, the output is
triggered using an internal trigger signal. Type in:
TT0 to disable the internal trigger
TT1 to enable the internal trigger using the period last set
TT1,p to enable the internal trigger and set the period.
Where:
p = trigger period in milliseconds (10ms to 5s).
Note 1: The TT1 command does not disable external triggers.
Note 2: This setting can be saved to non-volatile memory using the
AW command.
To view the status of the internal trigger, type in:
ST0
A typical reply to this command might be:
TM1,TP20.00ms
Where:
TM = Trigger status. 1 is on, 0 is o
TP = Trigger period
Showconguration
This command shows the conguration of the RC120. Type in:
ST