RC1xx LED Lighting Controllers - User Manual
33
A typical reply to this command might be:
CH1,MD0,S50.0,0.0,DL1.000ms,PU1.000ms,RT
0.0us,IP1,FL0,CS0.000A,RA0.000A
Where:
CH Channel number; this is always 1.
MD Mode: 0 = continuous, 1 = pulsed, 2 = switched.
S Percentage brightness: 50.0 = 50%; the second part is
always 0.0.
DL Pulse delay.
PU Pulse width.
RT Re-trigger delay.
IP Always return IP1.
FL Flags set using the RE1 command.
CS Current rating of the light after being sensed
successfully by SafeSense.
RA Congured voltage or current rating of the light set by
the VL command.
Clearconguration
CL
This command clears the conguration and lighting ratings, and turns
the output o. The results of any VL, RU, RP, RS, RW, RT, RE, TT, and
AW commands are cleared.
Simulate an input trigger
TR1
This command simulates an input trigger. If the RC120 is in pulse
mode, it pulses the output once.
Enable Ethernet messages
This command enables or disables Ethernet messages. Type in:
GTm
Where:
m = 0 to disable Ethernet messages, and
m = 1 to enable Ethernet messages.