Page 48 of 53  Copyright © 2007, General Electric Company  2023486-201 Rev C 
 
 
Guidance and manufacturer’s declaration – electromagnetic emissions 
The AED PRO is intended for use in the electromagnetic environment specified below.  The customer or the user of the 
AED PRO should assure that it is used in such an environment. 
Emissions test  Compliance  Electromagnetic environment – guidance 
RF emissions 
 
CISPR 11 
Group 1 
The AED PRO uses RF energy only for its 
internal function.  Therefore its RF emissions 
are very low and are not likely to cause any 
interference in nearby electronic equipment. 
RF emissions 
 
CISPR 11 
Class B 
Harmonic emissions 
 
IEC 61000-3-2 
Not applicable 
Voltage fluctuations/flicker 
emissions 
 
IEC 61000-3-3 
Not applicable 
The AED PRO is suitable for use in all 
establishments, including domestic 
establishments and those directly connected 
to the public low-voltage power supply network 
that supplies buildings used for domestic 
purposes. 
 
Guidance and manufacturer’s declaration – electromagnetic immunity 
The AED is intended for use in the electromagnetic environment specified below.  The customer or the user of the AED 
should assure that it is used in such an environment. 
Immunity test  IEC 60601 test level  Compliance level 
Electromagnetic environment - 
guidance 
Electrostatic discharge 
(ESD) 
 
IEC 61000-4-2 
±6 kV contact 
 
±8 kV air 
±6 kV contact 
 
±8 kV air 
Floors should be wood, concrete or 
ceramic tile.  If floors are covered 
with synthetic material, the relative 
humidity should be at least 30% 
Electrical fast 
transient/burst 
 
IEC 61000-4-4 
±2 kV for power supply 
lines 
 
±1 kV for input/output 
lines 
Not applicable   
Surge 
 
IEC 61000-4-5 
±1 kV differential mode 
 
±2 kV common mode 
Not applicable   
Voltage dips, short 
interruptions and 
voltage variations on 
power supply input 
lines 
 
61000-4-11 
<5 % U
T
 
(>95 % dip in U
T
) for 0.5 
cycle 
 
40 % U
T
 
(60 % dip in U
T
) 
for 5 cycles 
 
70 % U
T
 
(30 % dip in U
T
) 
for 25 cycles 
 
<5% U
T
 
(>95% dip in U
T
) 
for 5 sec 
Not applicable   
Power frequency 
(50/60 Hz) magnetic 
field 
 
IEC 61000-4-8 
3 A/m  80 A/m  Power frequency magnetic fields 
should be at levels no higher than 
those characteristic of a typical 
location in typical heavy industrial 
and power plants and the control 
rooms of H.V. sub-stations. 
NOTE   U
T
 is the a.c. mains voltage prior to application of the test level.