Section 7: Accessories 
2023486-201 Rev C  © 2007 General Electric Company  Page 49 of 53 
 
Guidance and manufacturer’s declaration – electromagnetic immunity 
The AED PRO is intended for use in the electromagnetic environment specified below.  The customer or the user of the 
AED PRO should assure that it is used in such an environment. 
Immunity test  IEC 60601 test level 
Compliance 
level 
Electromagnetic environment - guidance 
 
 
 
 
 
 
 
 
 
Conducted RF 
 
IEC 61000-4-6 
 
 
 
 
 
 
 
 
 
Radiated RF 
 
IEC 61000-4-3 
 
 
 
 
 
 
 
 
 
3 Vrms 
 
150 kHz to 80 MHz outside 
ISM bands
a
 
 
 
10 Vrms 
 
150 kHz to 80 MHz 
in ISM bands
a
 
 
 
10 V/m 
 
80 MHz to 2.5 GHz 
 
 
 
 
 
 
 
 
 
Not Applicable 
 
 
 
 
 
 
Not Applicable 
 
 
 
 
 
 
10 V/m 
Portable and mobile RF communications 
equipment should be used no closer to any part of 
the AED, including cables, than the recommended 
separation distance calculated from the equation 
applicable to the frequency of the transmitter. 
 
Recommended separation distance 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
d = 1.2 √P  80 MHz to 800 MHz 
 
d = 2.3 √P  800 MHz to 2.5 GHz 
 
 
where P is the maximum output power rating of the 
transmitter in watts (W) according to the 
transmitter manufacturer and d is the 
recommended separation distance in meters (m)
b
. 
 
Field strengths from fixed RF transmitters, as 
determined by an electromagnetic site survey,
c
 
should be less than the compliance level in each 
frequency range.
d
 
 
Interference may occur in the vicinity of equipment 
marked with the following symbol: 
 
 
NOTE 1  At 80 MHz and 800 MHz, the higher frequency range applies. 
 
NOTE 2  These guidelines may not apply in all situations.  Electromagnetic propagation is affected by absorption and 
reflection from structures, objects and people.