7-16 ApexPro™ 2001989-200E
Viewing real-time patient data
Pace detection switches to Single-Lead when using a 3-leadwire set.
For more information, refer to “Pacemaker troubleshooting” on page D-3.
Pace help
Clicking on the Pace Help button opens a window that shows common problems and
solutions in regard to pacemaker detection. This window is shown below.
047A
Lead analysis
The Lead Analysis control signals the transmitter to process the ECG in Single-Lead
or Multi-Lead mode. Use the mouse to click on your selection. Multi-Lead analysis is
the default setting for Adult.
NOTE
ECG is relearned whenever Lead Analysis is changed.
Multi-Lead analysis
Multi-Lead analysis simultaneously examines ECG leads I, II, III, and V (whether
they are displayed or not) to help eliminate false alarms and improve the ability of the
system to:
Detect beats which occur isoelectric to a single chest lead.
Discriminate artifact that appears in one lead compared to the other lead vectors.
Provide a smart-lead fail feature, where the failed lead is identified, and if
available, another lead is provided for display.
Continue arrhythmia processing even after a lead change.
Single-Lead analysis
Single-Lead analysis uses only the lead displayed on the CIC Pro center screen to
process ECG and arrhythmia information. To change the lead used for Single-Lead
analysis, you must change the displayed lead.