4-18 ApexPro™ 2001989-200E
System setup
For detailed control setting information, including displayed waveform, Alarm
Control, parameter control and print settings, refer to the CIC Pro Clinical
Information Center Operators Manual.To adjust the Alarm Volume, refer to Alarm
volume on page 4-18.
Alarm volume
To adjust the Alarm Volume for a telemetry patient at the CIC Pro center, follow these
steps:
1. From the multi-patient viewer, click CIC Setup > CIC Defaults to display the
current alarm volume.
053A
2. Click the down arrow next to Current and select the desired volume setting. The
current alarm level determines the actual alarm volume.
NOTE
The Minimum Alarm Volume is set between OFF and 100% in the service
mode. The Current Alarm Volume cannot be set below the Minimum
Alarm Volume.
3. Click the down arrow next to Minimum and select the desired Alarm Volume.
The minimum Alarm Volume can be set between OFF and 100%.
4. Click Apply.