E-4 ApexPro™ 2001989-200E
Technical specifications
Physical specifications
FCC compliance information
This device complies with Part 95 of the FCC Rules and RSS-210 of Industry
Canada.
Operation of this equipment requires the prior coordination with a frequency
coordinator designated by the FCC for the Wireless Medical Telemetry Service.
Certifications
UL/IEC/EN 60601-1
CAN/CSA C22.2 No.601.1
IEC/EN 60601-1-1
IEC/EN 60601-1-2
IEC/EN 60601-1-4
IEC/EN 60601-2-27
IEC/EN 60601-2-49
CE marked to the Medical Devices Directive 93/42/EEC
608.025 to 613.975 MHz - FCC Part 95 (ApexPro CH)
ECG gain accuracy ± 5% @ 15 Hz
Common mode rejection 100 dB minimum @ 60 Hz
Defibrillator protection ± 5000 VDC, 360 joules into 100 ohm
Defibrillator recovery time Defibrillator recovery time recovers within 5 seconds
Pacemaker detection ApexPro: ± 2 mV to ± 700 mV (RTI); 100 µsec to 2 msec;
either polarity
ApexPro CH: ± 2 mV to ± 700 mV (RTI); 100 µsec to 2 msec;
either polarity; on multiple leads
Patient leakage current Meets UL/IEC 60601-1
Serial communications 2 ports at 9600 baud asynchronous
Height 13.7 cm (5.38 in)
Width 7.3 cm (2.875 in)
Depth 2.3 cm (0.91 in)
Weight 141.8 g (0.275 lb) without battery; 170.1 g (0.375 lb) with
battery