System setup
2001989-200E ApexPro™ 4-11
NOTE
The Telemetry Unit Defaults tab sheet settings take precedence when the Patient
Age chosen in the Admit tab sheet and the age default setting from the Telemetry
Unit Defaults tab sheet match.
WHEN THE AGES DO NOT MATCH, THE Admit TAB SHEET AGE
SETTING TAKES PRECEDENCE.
The options found in the pull-down list for the Patient Age field are:
0-2 years
3-10 years
11-13 years
Adult (factory default selection)
The Patient Age setting chosen in the Telemetry Unit Defaults tab sheet affects the
alarm settings.
Factory defaults
These factory defaults are in effect, depending upon the patient’s age, unless they
have been modified through Telemetry Unit Defaults.
ECG Display Lead is II
Multi-Lead analysis
Telemetry Unit Defaults
Tab Sheet Setting
Age Chosen In The Admit Tab
Sheet When Patient Admitted
Resulting Limits
Unit Default
Alarm Level (Brady)
Resulting
Alarm Level
Adult Adult
0-2 years
3-10 years
11-13 years
50, 150
90, 200
60, 180
50, 150
Advisory
Advisory
Advisory
Advisory
Advisory
Crisis
Crisis
Advisory
0-2 years Adult
0-2 years
3-10 years
11-13 years
50, 150
90, 200
60, 180
50, 150
Advisory
Advisory
Advisory
Advisory
Advisory
Advisory
Crisis
Advisory
3-10 years Adult
0-2 years
3-10 years
11-13 years
50, 150
90, 200
60, 180
50, 150
Advisory
Advisory
Advisory
Advisory
Advisory
Crisis
Advisory
Advisory
11-13 years Adult
0-2 years
3-10 years
11-13 years
50, 150
90, 200
60, 180
50, 150
Advisory
Advisory
Advisory
Advisory
Advisory
Crisis
Crisis
Advisory