4-14 ApexPro™ 2001989-200E
System setup
Full disclosure defaults
This option sets the full disclosure settings. In user mode, only the full disclosure
Report and Strip settings can be configured. You must be in the Service mode to set
the other full disclosure settings at the CIC Pro center.
RR-APNEA seconds 30 Warning
PVC #/min 6 Advisory
Arrhythmia Alarm
Levels
Levels
ASYSTOLE
1
1
The default alarm level for Asystole and VFIB/VTACH cannot be moved from the
Crisis
level.
Crisis
VFIB/VTAC Crisis
V TACH Crisis
VT > 2 Crisis
V BRADY Crisis
ACC VENT Advisory
PAUSE Advisory
TACHY Advisory
BRADY Advisory
R ON T Message
COUPLET Message
BIGEMINY Message
TRIGEMINY Message
PVC Message
IRREGULAR Message
ATRIAL FIB Message
System Alarm Levels Levels
CHANGE BATTERY System Warning
OFF NETWORK System Warning
ARR SUSPEND System Warning
LEADS FAIL System Warning
PROBE OFF System Warning
Parameter Limits and Alarm Levels Low High Level