1-4 Installation
CE Mark Compliance For CE Mark compliance, the AT868 must meet both the EMC and
LVD directives.
IMPORTANT: CE Mark compliance is required for all units intended
for use in EEC countries.
EMC Compliance For EMC compliance, the electrical connections must be shielded and
grounded as in Table 1-1 below. Also refer to Figure 1-1 on the next
page an example of acceptable wiring. After all the necessary
electrical connections have been made, seal any unused cable entry
holes with standard conduit plugs or equivalent.
Note: If the instructions in this section are followed, the unit will
comply with the EMC Directive 89/336/EEC.
Table 1-1: Wiring Modifications
Connection Cable Type
Termination
Modification
Transducer RG62 a/u Add metallic cable clamp
from braid to chassis
ground.
Armored RG62 a/u or
conduit
Terminate RG62 a/u
shield to chassis ground.
Input/Output 22 AWG Shield
(e.g. Baystate #78-1197)
Terminate shield to
chassis ground at the
AT868.
Armored conduit None - grounded via
cable gland.
Power 14 AWG, 3 conductor,
shielded
(e.g. Belden #19364)
Terminate shield to
chassis ground at the
AT868.
Armored Conduit None - grounded via
cable gland.
Shielding Wires enclosed in a properly-grounded metal
conduit do not require additional shielding.