Programming Site Data 2-9
September 2004
Fluid Type 1. The selections for fluid type vary depending on whether the
Tracking Window is enabled or disabled. Press [
W] and [X] to
select the desired fluid and press
[ENTER]. Refer to Table 2-2
below for a list of available fluids.
2. If you selected
OTHER, enter the additional information required
by using the arrow keys. Press [
W] and [X] to move the cursor to
the desired location and press
[S] and [T] to increase or
decrease the number. Then press
[ENTER].
Reynolds Correction 1. Using [W] and [X], select Activ to activate the Reynolds
Correction Factor or select
Off to deactivate the Reynolds
Correction Factor. Then press
[ENTER].
Note: usually, Reynolds Correction should be set to
Activ.
2. Do one of the following:
If you selected OFF, enter the Calibration Factor and press
[ENTER]. Then, proceed to one of the following sections:
Clamp-on Transducers - proceed to Number of Traverses
and Transducer Spacing on the next page.
Wetted Transducers - The meter returns to the
Channel PROGRAM window. To complete initial setup of
the AT868, you must enter data in the
GLOBL-SYSTM
submenu on page 2-12.
If you selected ACTIV, proceed to KV Input Selection on the next
page.
Table 2-2: Fluid Types
Tracking Window = NO Tracking Window = YES
WATER W100 (Water, 0°-100°C)
OTHER W260 (Water, 0°-260°C)
Oil, Tracking
OTHER