Programming Site Data 2-17
September 2004
Setting Up a Totalizer
Output
1. At Output B or D, select OFF or TTLZR and press [ENTER].
2. Do one of the following:
If you selected OFF, the meter returns to the Global I/O menu.
Go to Configuring the Communications Port on the next page.
For 1-channel meter, go to step 4.
For a 2-channel/path meter, go to step 3.
3. At
Channel, press [X] to select the desired channel and press
[ENTER].
4. Press [
X] to select the desired measurement parameter and press
[ENTER].
5. Enter a value for the minimum pulse on-time (between 50 μsec
and 500,000 μsec) for the frequency of the totalizer pulses and
press
[ENTER]. Press [W] and [X] to move the cursor to the
desired location and press
[S] and [T] to increase or decrease
the number. Then press
[ENTER].
Note: A complete pulse consists of equal amounts of
ON and OFF
times. Choose a value that is compatible with the pulse
counter to be used.
6. At
UNITS/PULSE enter a value for the number of measurement
units represented by each pulse and press
[ENTER].