Programming Site Data 2-5
September 2004
Special Transducers 1. For a special transducer, press [X] to move the cursor to STD and
press
[S] until SPEC appears and press [ENTER].
2. Assign a number between 90 and 99 to the special transducer.
Enter the number by using the arrow keys. Press [
W] and [X] to
move the cursor to the desired location and press
[S] and [T] to
increase or decrease the number. Then press
[ENTER].
3. Press [
W] and [X] to select the wedge type (supplied by GE) and
press
[ENTER].
4. Press [
W] and [X] to select the transducer frequency (supplied by
GE) and press
[ENTER].
5. Enter the special transducer time delay value (supplied by GE) by
using the arrow keys. Press [
W] and [X] to move the cursor to
desired location and press
[S] and [T] to increase or decrease
the number and press
[ENTER].
Special Clamp-On
Transducers
The following two prompts appear only if special clamp-on
transducers are being used. If special wetted transducers are being
used, proceed to Pipe Outside Diameter on page 2-7.
1. Enter the wedge angle (supplied by GE) of the transducer by using
the arrow keys. Press [
W] and [X] to move the cursor to the
desired location and press
[S] and [T] to increase or decrease
the number. Then press
[ENTER].
2. Enter the wedge sound speed (supplied by GE) of the transducer
by using the arrow keys. Press [
W] and [X] to move the cursor to
the desired location and press
[S] and [T] to increase or
decrease the number. Then press
[ENTER] and proceed to Pipe
Material on the next page.