September 2004
2-16 Programming Site Data
Setting Up a Frequency
Output
1. At Output B or D, select OFF or FREQ and press [ENTER].
2. Do one of the following:
• If you selected OFF, the meter returns to the Global I/O menu.
Go to Setting Up a Totalizer Output on the next page.
• For 1-channel meter, go to step 4.
• For a 2-channel meter, go to step 3.
3. At
Channel, press [X] to select the desired channel and press
[ENTER].
4. Press [
X] to select the desired measurement parameter and press
[ENTER].
Note: The measurement units that appear in these prompts are those
selected in the
GLOBL-SYSTM menu earlier in this chapter.
5. At
BASE, enter a value for the low end of the output range
(-2,000,000 to 2,000,000) by using the arrow keys. Press [
W] and
[
X] to move the cursor to the desired location and press [S] and
[T] to increase or decrease the number. Then press [ENTER].
6. At
FULL, enter a value for the high end of the output range
(-2,000,000 to 2,000,000) by using the arrow keys. Press [
W] and
[
X] to move the cursor to the desired location and press [S] and
[T] to increase or decrease the number. Then press [ENTER].
7. At FULL SCALE FREQ, enter a value for the full scale of the output
(between 10 to 100,000) by using the arrow keys. Press [
W] and
[
X] to move the cursor to the desired location and press [S] and
[T] to increase or decrease the number. Then press [ENTER].
8. At Error handling, press [X] to select the desired option for
handling errors (
HOLD, LOW, HIGH, or OTHER) and press [ENTER].
9. If
OTHER is selected, the Error frequency display appears. This
allows the user to enter a set value for this option. Enter the
desired error frequency by using the arrow keys. Press [
W] and
[
X] to move the cursor to the desired location and press [S] and
[T] to increase or decrease the number. Then press [ENTER].