September 2004
1-10 Installation
Wiring the
Totalizer/Frequency
Output
The AT868 also provides a second output for each channel
(designated as Output B and Output D) that can be configured as a
totalizer or frequency output using the Instrument Data Manager
(IDM) software. These outputs can be configured independently.
Typically, Output B is used for Channel 1 and Output D is used for
Channel 2. However, either output can be configured for use with
either channel.
Figure 1-2 below shows a wiring diagram for a typical totalizer or
frequency output circuit.
Figure 1-2: Totalizer/Frequency Output Wiring
!WARNING!
NEVER CONNECT EXTERNAL POWER TO THE
PULSE/FREQUENCY OUTPUT TERMINALS.
To wire the totalizer/frequency outputs, complete the following steps:
1. Refer to Figure 1-5 on page 1-16 for the location of the terminal
blocks and wire the totalizer/frequency outputs as shown. Then,
secure the cable clamp.
Note: Outputs 1 and 2 in the wiring diagram correspond to Outputs
B and D in the AT868 software.
2. Do one of the following:
• Proceed to another section to continue wiring the AT868.
• Reinstall the plastic shroud, replace the front cover on the
enclosure and tighten the screws.
+5V
200
OUT
RTN
IN
Common
Model AT868 Pulse/Frequency Counter