Programming Site Data 2-15
September 2004
Setting Up the Analog
Output Scale (cont.)
6. At FULL, enter a flow rate value for the high end of the analog
output range by using the arrow keys. Press [
W] and [X] to move
the cursor to the desired location and press
[S] and [T] to
increase or decrease the number. Then press
[ENTER].
7. At
Error Handling, press [X] to select the desired option for
handling errors (
HOLD, LOW, HIGH, or OTHER) and press [ENTER].
8. If
OTHER is selected, the Forced milliamps display appears. This
allows the user to impose a set value for this option. Enter the
desired forced milliamps value by using the arrow keys. Press [
W]
and [X] to move the cursor to the desired location and press [S]
and [T] to increase or decrease the number. Then press [ENTER].
Accessing the
Totalizer/Frequency
Outputs
A totalizer output issues one pulse per selected volume of flow, and
the meter produces a pulse each time the programmed amount of flow
passes through the pipe. A frequency output issues a frequency that is
proportional to the assigned measurement parameter.
Output B and Output D are totalizer/frequency outputs that can be
configured independently. To configure these outputs, complete the
following steps:
1. Press
[ESC], [ENTER], [ESC] in sequence to enter the user program.
2. Press [
X] until PROG appears and press [ENTER].
3. Press [
X] until GLOBL appears and press [ENTER].
4. Press [
X] until I/O appears and press [ENTER].
5. Press [
X] until OPTN appears and press [ENTER].
6. Press [
X] to select B or D and press [ENTER].
7. Do one of the following:
• Frequency Output - go to Setting Up a Frequency Output on
the next page.
• Totalizer Output - go to Setting Up a Totalizer Output on
page 2-17.