2-12 Programming Site Data
Entering Global System
Data [GLOBL-SYSTM
Menu]
Refer to Figure A-2 on page A-2, and complete the steps below to
enter system information in the
GLOBL menu.
Selecting the
GLOBL-SYSTM Units
1. Press [ESC], [ENTER], [ESC] in sequence to enter the user program.
2. Press [
X] until PROG appears and press [ENTER].
3. Press [
X] until GLOBL appears and press [ENTER].
4. Press [
X] until SYSTM appears and press [ENTER].
5. Enter a short message (up to 16 characters) by using the arrow
keys. Press [
W] and [X] to move the cursor to the desired location
and press
[S] and [T] to select the desired character. Then press
[ENTER].
Selecting the System Units 1. Press [X] to select the system units and press [ENTER].
2. Do one of the following:
1-Channel Meters - go to Setting Up the External Totalizer
Reset Switch on the next page.
2-Channel Meters - continue with the next step.
3. Press [
X] to select the volumetric units for the flow rate display
and press
[ENTER].
4. Press [
X] to select the number of digits to the right of the decimal
point in the volumetric flow rate display and press
[ENTER].
5. Press [
X] to select the desired units for the totalized flow rate
display and press
[ENTER].
6. Press [
X] to select the desired number of digits to the right of the
decimal point in the totalized flow rate display and press
[ENTER].