Section 15 - Calibrate Channels
37
15. Calibrate Channels
Before you begin Channel Calibration, you must choose the full scale range and set the
appropriate jumpers (See Section 18.6). Calibrate both channels of your monitor. To
calibrate channels you will need these instruments and tools
• multimeter
• function generator
• screwdriver
Use this procedure to calibrate both channels.
Calibrating a channel involves these main steps:
• Set the filter option to No Filters
• Connect and adjust the test instruments
• Calibrate the channels
• Prepare to bring the monitor on line