3300/25 Dual Accelerometer Input Monitor Operation Manual
60
Channel B Integrator/Gain Stage Location
CHANNEL B
Main Board Filter Board
INTEGRATOR/GAIN STAGE
LOCATION
INSTALL REMOVE INSTALL REMOVE
*Integrator/Gain Stage with
NO Filtering
W5B,3B W3A,3C
W41,60
W61,85
W86
**Filter after the
Integrator/Gain Stage
W3B
W3A,3C
W5B
W41,60
W61,85
W86
Filter before the
Integrator/Gain Stage
W3A,3C
W5B
W3B W61
W41,60
W85,86
High Pass Filter before the
Integrator/Gain Stage
and Low Pass Filter after the
Integrator/Gain Stage
W3A,3C W3B,5B
W58,60
W85,86
W20,21
W41,59
W61
* Denotes standard configuration.
** When filtering without integration, set the filter after the Integrator/Gain Stage.
NOTE: If you have chosen the NO Filtering option, then you have completed your filter
Configuration.
When measuring velocity, low frequency signals can cause the monitor to saturate.