Maintenance
VisualInspection
Inspecttheacquisitionmoduleeachtimeyoucleanitorifyoususpectaproblem.
Checktheleadwiresandleadwireadaptersforwearandlooseconnections.
Replacethesepartsattherstsignofwear.
Checkthepinsthattheleadwirespluginto.Theyshouldnotbebentorloose.
ContactGEHealthcareCustomerSupportforanyrepairsneeded.
Checktheacquisitionmoduleplasticcaseforanydamage.ContactGEHealthcare
CustomerSupportforanyrepairsneeded.
Cleaning
Theexteriorandleadwiresoftheacquisitionmoduleshouldbecleaneddaily.
CleaningandDisinfectingAcquisitionModules
Propercleaninganddisinfectingprolongsthelifeofacquisitiondevices.Failuretouse
thepropercleaningsolutionsortofollowproperprocedurescanresultinthefollowing:
Appearanceofwaveformswhennotconnectedtoapatient,resultinginfalse
alarmsinsteadofleadfailurealarms
Brittleandcrackeddevicecase
Melting,dulling,ordistortionofthecase
Totaldevicefailure,requiringreplacement
Unitmalfunction
Voidedwarranty
Usethefollowingproceduretocleananddisinfectthecablesandleadwires.
1.Removecables,leadwires,andbatteriesfromthedevicebeforecleaning.
Makesuretormlyclosethebatterydoorafterremovingthebatteries.
2.Toclean,wipewithalightlymoistenedcloth.
Useamildsoapandwatersolutiontomoistenthecloth.
DoNOTuseanyofthefollowingcleaningproducts,orproductsthatcontainthe
sameactiveingredientsandsolutions,whichareknowntocausetheproblems
previouslylisted:
Sani-Cloth®Wipes
Ascepti®Wipes
HBQuat®
Clorox®Wipes(theydonotcontainbleach)
Over-the-counterdetergents(suchasFantastic®,Tilex®,andsoon)
14
CAM-14
®
/CAM-HD
®
2049751-001C