Troubleshooting
Troubleshooting
Symptom
Troubleshooting
Thehostdevicedisplaysthemessage
AcquisitionModuleDisconnected
Checktheacquisitioncableconnection.
Checkthehostcableinterfaceexinthe
module.
Theleadtracingsappearinvertedoratthe
wrongamplitude
Checkthepositionoftheleadwiresonthe
acquisitionmoduleandthepatient.
Allleadshaveexcessivenoise
ExaminetheRAandRLleadwiresand
replacethemifnecessary.
Checktheleadpreparationsite.
RefertoEinthoven'sTriangletodetermine
whichleadorleadsarecausingthenoise.
Movetheacquisitionmoduletoanarea
wherethereislesselectricalnoise.
Tryanotheracquisitionmodule.
Controlswitchbuttonsontheacquisition
modulefailorareintermittent
CheckthattheSwitchFlexCableis
pluggedintothePCB.
Replacetheswitchlabel.
Ifthereisstillafailure,replacethemain
PCB.
Einthoven’sTriangle
ThefollowingillustrationshowsEinthoven'sTriangleandthedenitionsofthe
standard12leads.Somendthisinformationusefulwhentroubleshootingnoisy
leadsorleadgroups.