Troubleshooting
ThischapteraddressesproblemswithECGdataacquisitionandthequalityofECG
data.Thetroubleshootinginformationinthischapterwillhelpyounarrowproblemsto
oneoftheeldreplaceableassemblies.
NOTE:
Acopyofagood12-leadreportgeneratedfromanECGsimulatorisveryuseful
whentryingtodetermineifthedataiscorrectonareport.
GeneralInformation
Thefollowingsectionsprovidegeneralinformationtohelpidentifyproblemswith
theacquisitionmodule.
ECGDataNoise
IftheacquiredECGdatadisplaysunacceptablenoiselevels:
Verifyproperelectrodeplacement.
Verifyproperelectrodeapplication.
(Perspirationanddeadskinmustberemovedfromtheelectrodesite.)
Checkfordefectiveorout-of-dateelectrodes.
Checkfordefective,broken,ordisconnectedleadwires.
Checkthepatient’sposition.
ThepatientshouldremainmotionlessduringtheacquisitionofarestingECG.
TypicalProblemsandSolutions
Thefollowingtableidentiessometypicalacquisitionmoduleproblemsandsolutions:
2049751-001C
CAM-14
®
/CAM-HD
®
23