ReassemblingtheAcquisitionModule
Strictantistaticprecautionsshouldbefollowedduringdisassembly/assemblyofthe
acquisitionmodule.
1.Applysiliconesealanttotheedge(1)ofthetopandbottomcovers.
2.PlacethemainPCBinthebottomcover.
3.Connectthehostcableinterfaceex(2)atthemainPCBbygentlypressing
theconnectorinplace.
4.Connecttheswitchexcable(3)tothehostcableinterfaceexatthemainPCB
(4)bygentlypressingtheconnectorinplace.
5.Whileholdingthetop(5)andbottomcoverstogether,replacethefourTorx
screws(6)usinga#6Torxdriver.DONOTovertighten.
6.PerformaFunctionalCheckoutprocedure.